Setra-system Model Datum 2000 User Manual

Browse online or download User Manual for Sensors Setra-system Model Datum 2000. Setra System Model Datum 2000 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - 2000™

1DATUM 2000™DATUM 2000™Dual Channel Meter and ManometerOperating Manual

Page 2 - Contents

10Step 5.Press the HOLD/mode key to display the number “0”.Note: The decimal point that was set during calibration (CAL_U, CAL_C) willbe displayed.S

Page 3

110.00 CALSCANHOLDadj modeSCAN HOLDadjmodenoYES CH1HI /LO ALARM SETPOINTS (please read before attemptingprocedure)Note: The channel must be calibrate

Page 4

12Step 5.Set the LO ALARM Point.Note: The decimal point that was set during calibration (CAL_U, CAL_C) willbe displayed.Press the SCAN/adj key to mov

Page 5 - Figure A

13INSTALLATION OF PRESSURE FITTINGS FORTHE MANOMETERYour transducer is designed for most accurate operation when subjected topressures within the desi

Page 6 - Terminal Block Connector

14CALIBRATING THE DATUM 2000™.Note: If on power up, the Datum 2000™ display alternates between "cal"and "no Yes", it has not bee

Page 7 - OPERATING NOTES

15CALIBRATING THE DATUM 2000™ USING AN EXTERNAL REFERENCEVOLTAGE:Connect the external reference to the meter to simulate the transducer ortransmitter

Page 8 - Cal is lit when

16PRESSURE (@ 0) = 200 x 0 + 0 = 0PRESSURE (@ REF) = 200 x 2.4995 + 0 = 499.92nd example:A 20 psi transducer has an offset of .1 volts and will have a

Page 9

17For the first pressure point connect the "mA" to gound, pin 8 to pin 9 (CH2:pin 14 to pin 15). Enter the pressure that corresponds to ze

Page 10 - SCAN HOLD

18CALIBRATING VOLTAGE INPUT AND CURRENT INPUT(please read before attempting procedure)CAL_U and CAL_C commands allow you to calibrate the DATUM 2000™

Page 11 - 0.00 CAL

19Step 4.Press the SCAN/adj key to display “OFFST”.Step 5.To calibrate voltage in (transducer), press the SCAN/adj key to display“CAL U”. (If calibra

Page 12

2ContentsINTRODUCTION ... 3CONNECTING A TRANSDUCER TO THE DATU

Page 13 - THE MANOMETER

20Step 7b.Press the HOLD/mode key to accept the decimal point and the meter willdisplay the decimal point with zeroes.Step 8.Enter a calibration data

Page 14 - CALIBRATING THE DATUM 2000™

21Step 9.Repeat the same procedure as in Step 8 to enter the second calibration datapoint. (Enter a number between +31999 and -9999.)After the second

Page 16

23SETRA DATUM 2000™ RS232 SERIAL OPTION INTERFACEINSTRUCTIONSThe serial option for the Setra DATUM 2000™ has a 9 pin d-sub connectorfor connecting the

Page 17 - SLOPE = ————————————————————

242 32H Set the DATUM 2000™ to channel two and lock the meterso only channel two is displayed. It then transmits onereading for channel two. Channel

Page 18

25CHANGING THE BAUD RATE AND PARITYThe baud rate and parity of the DATUM 2000™ Serial Option can be set tothe following values:COMMAND ASCIIBAUD: 300

Page 19 - +31999 to -9999

26DATUM 2000™ Manometer Specificationsw/Models 204/204D w/Model 239 w/Model 270Type of PressureMeasurementGage Differential GageAbsolute Gage Absolute

Page 20 - 30.0000.000

27DATUM 2000™ Meter and Manometer SpecificationsDisplay DescriptionType Light emitting diodeColor R e dHeight of Digits 0.56 inchMaximum Readout - 999

Page 21

28PANEL MOUNT INSTRUCTIONS1. Slide the panel mount tab with screwinto slot on the side of the meter asshown.

Page 22

29Note: You may need to bendthe panel mount tabs to theirnew position after they are used.2. Slide the screw down until it contactsthe panel. Now ti

Page 23 - INSTRUCTIONS

3INTRODUCTIONThank you for purchasing a Setra DATUM 2000™ meter or manometer.Setra’s DATUM 2000™ is a compact dual channel meter with a 4 1/2 digitLED

Page 24

30"FCC WARNING"Warning: Changes or modifications to this unit not expressly approved bythe party responsible for compliance could void the

Page 25

31RETURN INSTRUCTIONSPlease contact Setra (800-257-3872 or 978-263-1400) before returning unit for repairto review information relative to your applic

Page 26

4CONNECTING A TRANSDUCER TO THE DATUM 2000™Note: If you purchased your DATUM 2000™ factory calibrated with a Setrapressure transducer or transmitter,

Page 27

5Figure ATRANSDUCERS ( 4 WIRE OR 4 TERMINAL DEVICES)Connect the + excitation to pin 5 (CH2: pin 11) to supply 24 volts to thetransducer. IF YOUR TRAN

Page 28 - PANEL MOUNT INSTRUCTIONS

6123456789101112131415Channel 1Inputs/ExcitationsChannel 2Inputs/ExcitationsAlarmSignalsFigure BTerminal Block ConnectorSetra SetraTransducer* Transmi

Page 29

7OPERATING NOTESOvEr AND UndEr MESSAGESIn addition to displaying the numeric pressure reading, the DATUM 2000™may also display "UndEr" and O

Page 30 - "FCC WARNING"

8USER SETUPSThe menu tree below represents the meter program structure. To enter themenu and select a channel, press and hold the SCAN/adj and HOLD/m

Page 31 - RETURN INSTRUCTIONS

9SET “OFFSET” (please read before attempting procedure)Note: The channel must be calibrated before the offset is adjusted.The ”OFFSET” command allows

Comments to this Manuals

No comments