Please Note: Series 540 / 541 / 542 are designed and manufactured in accordance with soundengineering practice as defined by the Pressure Equipment
* In the event of fire the end user must ensure that the system pressure is vented to a safearea.* The effects of decomposition of unstable fluids sho
High Temperature Pressure Measurement: Possible using a length of piping or a "syphon" toisolate the instrument from the hot media. Since
Compensated Temperature RangeModel Electrical Connector Temperature540 / 542 10-6 Bayonet Connector -65°F to +250°F (-54°C to +120°C)541 10-6 Bayonet
MAINTENANCERoutine Inspection: Not required except for periodic inspection of the cable and molding to ensurethat these are neither damaged nor softe
CAUTIONCARE MUST BE TAKEN NOT TO TOUCH THE PRESSURE SENSITIVEDIAPHRAGM WHILE CLEANING THE PRESSURE PORT. FAILURE TOOBSERVE THIS PRECAUTION CAN CAUSE
W A R N I N G * IF INSTRUMENT IS SUBJECTED TO +400°C PLUS Some instruments use Viton sealant. Above +400°C this material decomposes produ